-
The test was repeated with a total sample size of 297. The disinfection effect of the pulsed UV technology on two microbial indicator bacteria on different carriers at different temperatures, expressed as the average killing log value (Tables 1–2). When the set temperature was 20.0 ℃, the average carrier temperature was 20.4 ℃, ranging from 20.1 ℃ to 20.7 ℃; when the set temperature was 0.0 ℃, the average carrier temperature was 1.3 ℃, ranging from −0.8 ℃ to 2.0 ℃; when the set temperature was −20.0 ℃, the average carrier temperature was −16.7 ℃, ranging from −18.7 ℃ to −15.0 ℃.
Test strain Bacterial vector Setting temperature (℃) Pulsed UV irradiation disinfection average elimination killing log value 15 s 30 s 1 min 2 min Staphylococcus aureus Paper sheets 20.0 1.42 2.81 6.40 >6.32 0.0 1.30 2.36 6.40 >6.32 −20.0 1.27 1.86 5.80 5.29 Stainless steel sheet 20.0 2.60 5.08 >6.32 >6.32 0.0 2.41 3.27 >6.32 >6.32 −20.0 1.51 3.38 >6.32 >6.32 Cloth sheet 20.0 >6.32 >6.32 >6.32 >6.32 0.0 >6.32 >6.32 >6.32 >6.32 −20.0 5.50 >6.32 >6.32 >6.32 Notes: Negative control sterile growth. Positive control number: 1.00×106–4.20×106 CFU/sheet. The mean logarithmic value of the positive control was 6.32. >6.32 indicates sterile growth in the test group.
Abbreviations: UV=ultraviolet; CFU=colony forming units.Table 1. Killing effect of pulsed UV light on Staphylococcus aureus under different temperature conditions.
Test strain Bacterial vector Setting temperature (℃) Pulsed UV irradiation disinfection average elimination killing log value 5 s 10 s 15 s 20 s 30 s 1 min 2 min Escherichia
coliPaper sheets 20.0 1.61 2.86 3.63 4.42 5.84 >6.31 >6.31 0.0 2.04 2.38 3.14 3.23 4.02 >6.31 >6.31 −20.0 1.82 2.01 2.45 3.37 3.77 >6.31 >6.31 Stainless steel sheet 20.0 2.65 4.41 4.37 >6.31 >6.31 >6.31 >6.31 0.0 1.98 3.60 3.62 >6.31 >6.31 >6.31 >6.31 −20.0 1.70 2.98 3.92 >6.31 4.35 >6.31 >6.31 Cloth sheet 20.0 2.75 4.28 >6.31 >6.31 >6.31 >6.31 >6.31 0.0 1.55 4.37 6.69 >6.31 >6.31 >6.31 >6.31 −20.0 1.37 4.88 6.21 >6.31 >6.31 >6.31 >6.31 Notes: Negative control sterile growth. Positive control number: 1.00×106–3.83×106 CFU/sheet. The mean logarithmic value of the positive control was 6.31. >6.31 indicates sterile growth in the test group.
Abbreviations: UV=ultraviolet; CFU=colony forming units.Table 2. Killing effect of pulsed UV light on Escherichia coli under different temperature conditions.
To enhance the comparability of the analysis, the same irradiation time as the Staphylococcus aureus was selected in the experimental results of Escherichia coli. That is, four sets of data with irradiation times of 15 seconds, 30 seconds, 1 minute, and 2 minutes. The four influencing factors of temperature, irradiation time, carrier type and test species were used as groups, respectively, and the mean killing log values were calculated for each group. The differences in the disinfection effect of each influencing factor were compared using a non-parametric test for independent samples. The value of P less than 0.05 indicates that the difference is statistically significant (Table 3).
The effect of temperature on the disinfection effect of pulsed UV light is limited. A non-parametric test of disinfection effectiveness for independent samples with temperature as a group at P>0.05, which was not significant, concluded that the difference in disinfection effectiveness at each temperature was not statistically significant. No significant effect of temperature on the disinfection effect of pulsed UV was found (Table 3).
The irradiation time is the influencing factor of pulse UV disinfection effect. A non-parametric test of disinfection effect for independent samples with irradiation time as a group found a significant P
-value, P<0.05. Further comparisons between groups, the P-values were less than 0.05 between all other groups except between the 1-minute and the 2-minute groups. This means that the differences in disinfection effectiveness between the groups were considered statistically significant within the irradiation time of 15 seconds to 1 minute. The longer the irradiation time, the better the disinfection effectiveness was. The carrier type is an influential factor in the effectiveness of pulsed UV disinfection. A non-parametric test of disinfection effectiveness was performed on independent samples using carriers as groups and found P<0.05. Further comparisons between groups, the P-values were less than 0.05 between any two groups, which means that the differences in disinfection effect of each carrier were considered statistically significant.
Influencing Factors Median (Q1, Q3) Statistics P Setting temperature 20.0 ℃
0.0 ℃
20.0 ℃6.31 (5.27, 6.32)
6.31 (3.72, 6.32)
6.26 (3.81, 6.31)1.792 0.408 Irradiation time 15 s
30 s
1 min
2 min3.63 (2.41, 6.21)
5.46 (3.38, 6.31)
6.31 (6.31, 6.32)
6.31 (6.31, 6.32)28.667 0.000 Bacterial vector Paper sheets
Stainless steel sheet
Cloth sheet5.55 (2.54, 6.31)
6.31 (3.70, 6.32)
6.32 (6.31, 6.32)23.292 0.000 Test strain Staphylococcus aureus
Escherichia coli6.32 (2.93, 6.32)
6.31 (4.74, 6.31)5.465 0.000 Abbreviation: UV=ultraviolet. Table 3. Comparison of the effect of different influencing factors on the effectiveness of pulsed UV disinfection.
The test strain is an influential factor in the effectiveness of pulsed UV disinfection. A non-parametric test of disinfection effect for independent samples with test strains as groups found P<0.05, which concluded that the difference in disinfection effectiveness between the two test strains was statistically significant.
HTML
Citation: |